Product introduction
1) Electronic electric loading, high precision sensors, with a unique closed loop control system.
2) With the touch screen interface, the operator can directly select the hardness scale, and the force value can be changed automatically in the test parameters.
3) According to the standard hardness block, each hardness scale is divided into high, middle and low sections for automatic correction.
4) The test force can be automatically corrected by standard dynamometer.
5) Automatically apply the initial test force and the main test force.
6) Powerful data processing function, including: HR, HB, HV and other hardness system conversion, set the qualified range, automatic alarm over the limit, can input the tester, sample name and other information.
7) More sample and test information, measurement data USB disk saved in EXCEL format, easy to edit and process.
Technical parameters
Model |
HLR-150/45DX |
|
Superficial Rockwell initial test force |
3.0kgf(29.4N) |
|
Superficial Rockwell total test force15kgf(147.1N) 30kgf(249.2N) 45kgf(441.3N) |
||
HRA、HRB、HRC、HRD、HRE、HRF、HRG、HRH、HRK、 HRL、HRM、HRR、HRP、HRS、HRV |
||
Rockwell measurement range |
||
Rockwell test range |
HR15N:70-94 HR30N:42-86 HR45N:20-77 HR15TW:67-93 HR30TW:29-82 HR45TW:10-72 |
|
standard |
Conversion scale |
Rockwell, Brinell and Vickers |
200mm |
||
indenter560 * 220 * 800 mm |
||
Weight |
70kg |
|
ower supply |
AC220 + 5%, 50 ~ 60 hz |